ELECTROCARDIOGRAM (ECG) EQUIPMENT CALIBRATION TEST LABORATORY OF NURSES STIKES AL INSYIRAH NEW PEKANBARU
Keywords:
Electrocardiogram, ECG, Health, Calibration, HeartAbstract
Heart disease is one of the many deadly diseases. The death rate of people due to heart disease in Indonesia is
29.1% or 17.1 million%/year. To reduce the lack of death rates caused by heart disease, by checking heart
health using electrocardiogram modalities. The process of the heart's electrical activity is monitored and
displayed in a graph on the monitor screen. ECG is used not only in hospitals but also in nursing laboratories as
a clinical practice aid for nursing students. All medical devices must be calibrated regularly. This study shows
the ECG calibration process with the aim of getting the results of whether the ECG in the STIKES Al-Insyirah
Pekanbaru lab is suitable for use. The electrocardiogram used with the GE brand, type Mac 600. The method
used is a direct comparison method with an electrocardiogram simulator. The test results get a voltage level
calibration value with a tolerance of 5% with a value of 5.10, and 20 mm/mV. The value of the recording rate
has a tolerance of 100-104 mm. while the ECG signal calibration obtained measurement results of 10 mm and is
still within the tolerance limit.
Downloads
References
Arifin, Jaenal, and Amalia Norma. 2019. "Image Processing on the Ekg Signal".
Electrical Media 11 (1): 27–33.
Jiju, K., B. Sreekumari, and Niranjan D. Kbendate. 2011. “Indigenous Development of
Medical Electrical Safety Analyzer.” MeMeA 2011 - 2011 IEEE International
Symposium on Medical Measurements and Applications, Proceedings, no. May: 167–
https://doi.org/10.1109/MeMeA.2011.5966675.
Rifali, Myza, and Dessy Irmawati. 2019. “Intelligent System for Detecting
Electrocardiogram (ECG) Signals for Classification of Normal and Abnormal Hearts
Using Artificial Neural Networks (ANN).” Elinvo (Electronics, Informatics, and
Vocational Education) 4 (1): 49–55. https://doi. org/10. 21831/elinvo. v4i1. 28242.
Rpm, Measurement, Al Islami, Ria Adillah, and Triana Rahmawati. 2017. “May 2017
Final Project Seminar,” 1–9.
Saimi, Mulyatno. 2021. “Design ECG Simulator Digital To Analog Converter R-2R”.
Journal of Information and Computer Technology. 7(1):156–68. https://doi.
org/10.37012/jtik. v7i1. 531.
Septyani, Hanna, Syaifudin, and Hj. Andjar Pudji. 2018. “Thermohygrometer
Calibrator.” Thermohygrometer calibrator 1–9.
Setiawidayat, Patience, and Suci Imani Putri . 2016. “Discrete ECG data filtering for
determination of pqrst in one cycle”. Sentia 8:48–55.
Wati, Erna Kusuma. 2021. “Testing and Calibration of Medical Devices on
Electrocardiograph.” STRING (Research and Technological Innovation Writing Unit) 6
(1): 50. https://d